MS1 Conductivity Meter Installation and Maintenance Guide

Covers conductivity and TDS measurement, boiler water sampling, cooling and neutralization, sensor inspection, and routine maintenance.

1. Conductivity measurement

The instrument has an on/off button and a three-position range selector switch: 0-200 μS/cm, 0-2 mS/cm, and 0-20 mS/cm (1 mS/cm = 1000 μS/cm). An indicator lamp shows the range currently selected.

Ensure that the sample temperature is as close as possible to 25°C and does not exceed 45°C.

Switch on the instrument and use the range selector switch on the right to select the highest range: 0-20 mS/cm.

Immerse the sensor in the sample with the protective sleeve fitted. The immersion depth must not exceed half the length of the sensor. Wait 15 seconds for temperature compensation to take effect.

Read the display.

If a more accurate reading is required, switch to the next lower range.

If the current range cannot provide a reading, the display will show “1”.

When the display shows “1”, switch the range switch to the next higher range.

Example: sensor in tap water at 25°C:

Selected rangeReading× 1000 = μS/cm
20 mS/cm0.52 mS/cm520 μS/cm
2 mS/cm0.517 mS/cm517 μS/cm
200 μS/cm1N/A

After measurement, rinse the sensor and switch off the instrument; the instrument will also switch off automatically after approximately 5 minutes.

2. TDS measurement

The conductivity meter can be used to estimate the TDS value of a water sample. For neutral water samples, the temperature-compensated conductivity reading can be converted to TDS (ppm) using the following formula:

TDS (ppm) = conductivity reading (μS/cm) × 0.7

3. Boiler water sampling

Obtaining a representative boiler water sample is essential for analysis. The initial water taken from a water level gauge drain or an external water level control column is mainly condensate and is a relatively pure sample. This condensate must be drained before a true sample can be obtained.

Schematic diagram of the MS1 conductivity meter

A sample cooler should be used when sampling boiler water. If water is taken directly from the boiler, the pressure drop will cause some of the water to flash into steam and escape. This is very dangerous, and the flash losses will make subsequent analysis inaccurate. Using a sample cooler also brings the sample temperature close to 25°C and saves time.

4. Neutralizing boiler water samples

Acids and alkalis have a significant effect on conductivity. Therefore, before measuring the TDS of a boiler water sample, the alkalinity of the sample must first be neutralized.

The procedure is as follows:

Add a few drops of phenolphthalein indicator to the cooled sample; if the sample is alkaline, the solution will turn purple.

Add acetic acid (usually a 5% solution) drop by drop to neutralize the sample, stirring continuously until the purple color disappears.

Measure the conductivity and convert it to a TDS value as described in Section 2.

If neutralization cannot be carried out on site, the boiler TDS value can be estimated using the following formula:

TDS = conductivity reading (μS/cm) × 0.5

5. Maintenance

The sensor should be cleaned frequently according to use: remove the protective sleeve, gently clean the carbon electrodes with a brush and ordinary kitchen detergent, rinse thoroughly with tap water, and refit the protective sleeve.

Replace the battery when the display shows a low-battery warning.

Open the battery compartment on the back of the instrument to access a single 9V battery (PP3 or equivalent).

The instrument has no other serviceable parts. The sensor is permanently connected to the instrument; if it is damaged, the complete instrument must be returned for repair.

6. Using the extension lead

The instrument is supplied with an extension lead for measuring the AC resistance of an installed conductivity sensor in order to check its condition.

To use the extension lead, plug one end into the instrument and connect the other end directly to the sensor terminals using crocodile clips. Press the 20 mS/cm range switch. The AC resistance value (Ω) equals 100 divided by the displayed reading.

Example:

Displayed reading 5.00

Ω = 100 ÷ 5.00 = 20 Ω

The minimum measurable resistance is 5 Ω.

7. Determining the condition of the sensor

To determine the condition of the sensor, first determine its cell constant (which is related to the liquid conductivity and the sensor resistance).

To determine the cell constant:

Multiply the conductivity at 25°C by the factor in the table to obtain the conductivity at the boiler temperature.

Boiler pressure diagram

Boiler pressure (bar g)5710152032
Conductivity factor3.683.914.184.534.85.28
  1. Divide the conductivity reading (μS/cm) by 1000000 to convert it to S/cm.
  2. Multiply the resulting value by the sensor resistance (Ω, measured as described in Section 6) to obtain the cell constant.

Typical cell constants for sensors in BCS1, 2, or 4 systems are approximately 0.3; in BCS3 systems, approximately 0.3 to 0.7.

If the cell constant of a sensor in a BCS1, 2, or 4 system is greater than 0.7, or greater than 1.0 in a BCS3 system, the sensor may have failed and should be removed for inspection, cleaned, or replaced if necessary.

A low cell constant indicates that the sensor is working well; a high cell constant indicates that the conductivity at the sensor tip is already very low, possibly due to scaling.

An extremely low cell constant indicates an internal short circuit. The farther the sensor tip is from the boiler, the greater the cell constant.

Example:

A BCS3 system is installed on a boiler operating at 10 bar g (saturated steam temperature 184°C).

The conductivity meter reading for an unneutralized sample at 25°C is 4800 μS/cm.

The sensor resistance measured using the conductivity meter and extension lead is 20 Ω.

  1. Multiply the conductivity by the factor for a 10 bar boiler in the table above: 4800 × 4.18 = 20064
  2. Divide the result by 1000000 to convert it to S/cm: 20064 ÷ 1000000 = 0.020064
  3. Multiply the resulting value by the sensor resistance to obtain the cell constant: 0.020064 × 20 = 0.40128

The cell constant of this sensor is within the range of 0.3 to 0.7, so no further inspection is required.

8. Calibration

The instrument is calibrated at the factory and requires no adjustment during normal use.

If calibration with a standard solution is required or higher accuracy is desired, unscrew the plug on the left side of the display and adjust the potentiometer with a small screwdriver. The calibration adjustment range is 20 turns ±20% (approximate).

Calibration does not affect resistance measurements made with the extension lead.

See the product introduction, technical specifications, and supporting documentation for the MS1 conductivity meter.

FAQ

How do I measure sample conductivity correctly and select the appropriate range?

The sample temperature should be as close to 25°C as possible and must not exceed 45°C. After switching on the instrument, first select the highest range, 0-20 mS/cm. Immerse the sensor in the sample to a depth no greater than half the sensor length. Wait approximately 15 seconds for temperature compensation to take effect, then read the displayed value. For a more precise reading, switch to the next lower range. If the display shows "1", the current range cannot provide a reading, so switch to the next higher range. After measurement, clean the sensor and switch off the instrument.

How can I ensure that a boiler water sample is representative?

At the start of sampling, water taken from the water-level gauge drain or the external water-level control chamber is primarily condensate. This condensate should be discharged before the actual sample is collected. Use a sample cooler to collect a boiler water sample. When sampling directly, the pressure drop causes some water to flash off, which is both dangerous and likely to make subsequent analysis inaccurate. A sample cooler also brings the sample temperature close to 25°C.

How do I convert a conductivity reading to a TDS value?

For neutral water samples, convert the temperature-compensated conductivity reading using TDS (ppm) = conductivity reading (μS/cm) × 0.7. Before measuring boiler water TDS, neutralize the sample's alkalinity: add a few drops of phenolphthalein indicator. If the sample turns purple, add approximately 5% acetic acid drop by drop and mix until the purple color disappears, then take the measurement. If neutralization is not possible on site, estimate the boiler water TDS using TDS = conductivity reading (μS/cm) × 0.5.

What should I pay attention to during routine maintenance?

Clean the sensor regularly according to the usage conditions. Remove the protective sleeve, gently clean the carbon electrodes with a brush and ordinary kitchen detergent, rinse them thoroughly with tap water, and refit the protective sleeve. Replace the battery when the display shows a low-battery warning. The battery is a single 9V battery (PP3 or equivalent). The sensor is permanently wired to the instrument. If it is damaged, return the complete instrument for repair. The remaining components require no maintenance.

Does the instrument require calibration?

The instrument is calibrated before shipment and requires no adjustment during normal use. If calibration with a standard solution is required, unscrew the pin on the left side of the display and adjust the potentiometer with a small screwdriver. The calibration range is approximately 20 turns ±20%. Calibration does not apply when measuring resistance with an extension lead.

Documents

Recommended products

Our Contact Details

Click email or phone to copy. You can also scan WhatsApp / WeChat.

WhatsApp QR code WhatsApp
SteamVendor WeChat QR code WeChat